Operations 2 – Teledyne TED-60 - Medical application oxygen monitor User Manual

Page 11

Advertising
background image

Operations 2

Operations 2

Operations 2

Operations 2

Operations 2

2-3

Teledyne Analytical Instruments

Calibration

Calibration

Calibration

Calibration

Calibration

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Never expose the sensor to varying temperatures (i.e., hold the sensor in your

Never expose the sensor to varying temperatures (i.e., hold the sensor in your

Never expose the sensor to varying temperatures (i.e., hold the sensor in your

Never expose the sensor to varying temperatures (i.e., hold the sensor in your

Never expose the sensor to varying temperatures (i.e., hold the sensor in your
hand) while calibrating.

hand) while calibrating.

hand) while calibrating.

hand) while calibrating.

hand) while calibrating.

1. Switch the unit ON using “

0/I

” ON/OFF key.

2. Place the sensor in 100% oxygen.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: The flow diverter should only be used with flowing gas circuits.

The flow diverter should only be used with flowing gas circuits.

The flow diverter should only be used with flowing gas circuits.

The flow diverter should only be used with flowing gas circuits.

The flow diverter should only be used with flowing gas circuits.

3. The CALIBRATE button is located on the top of the unit. Turn the button

by pressing down on it and twisting. Rotate until the display reads 100%.

4. Remove the sensor from 100% oxygen. If the flow diverter is attached,

remove it and expose the sensor to room air. The LCD should display 20–
22%.

5. A calibration check may be conducted at any time to assure proper opera-

tion. Remove the diverter and expose the sensor to room air, and verify that
the reading is 20–22%. Expose the sensor to 100% oxygen (preferably using
the diverter and tee adapter in a flow circuit) and verify that the reading is
99–101%.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Never calibrate the unit in humidified gas, as water vapor dilutes the calibration

Never calibrate the unit in humidified gas, as water vapor dilutes the calibration

Never calibrate the unit in humidified gas, as water vapor dilutes the calibration

Never calibrate the unit in humidified gas, as water vapor dilutes the calibration

Never calibrate the unit in humidified gas, as water vapor dilutes the calibration
gas and makes the oxygen concentration appear lower than it really is.

gas and makes the oxygen concentration appear lower than it really is.

gas and makes the oxygen concentration appear lower than it really is.

gas and makes the oxygen concentration appear lower than it really is.

gas and makes the oxygen concentration appear lower than it really is.

Use

Use

Use

Use

Use

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Prior to use, check the calibration and check the sensor for leaks or damage.

Prior to use, check the calibration and check the sensor for leaks or damage.

Prior to use, check the calibration and check the sensor for leaks or damage.

Prior to use, check the calibration and check the sensor for leaks or damage.

Prior to use, check the calibration and check the sensor for leaks or damage.

The TED 60T can be used to analyze gas mixtures for oxygen in two basic modes:

1.) In breathing circuits or other instances where gases are flowing in tubing

circuits.

2.) In confined volumes such as incubators or tents.

When measuring for oxygen in breathing circuits, the diverter must

must

must

must

must be used. The

diverter should be screwed onto the threaded front end of the T-7 sensor. A tee adapter

Advertising