Accessories

Description Catalog

No.

Rotameter
Measures 4 to 50 L/min on a 4-inch scale
at 3% accuracy (full scale)

320-440

Certifi cate of Compliance, optional,
must be ordered when ordering pump P1501

Tubing, polyurethane, 1/4-in ID, 15/32-in OD,
reinforced to prevent kinking, 10 feet

225-1350

4. Sample for the desired sampling period.
5. Turn

off pump and unplug from wall outlet.

6. Package sample appropriately for shipment to a laboratory (see BioSampler Operat-

ing Instructions).

Note: Sonic fl ow is att ained when a vacuum pump pulls air through an orifi ce to create a condi-
tion where the downstream pressure is 52% of the upstream pressure. At this point, air velocity
through the orifi ce has reached the speed of sound and cannot increase, therefore, the orifi ce
becomes “critical.” The fi xed air velocity at the critical orifi ce also limits the airfl ow so that the
condition remains constant regardless of pressure variations in the system. Airfl ow will remain
constant at vacuum levels ≥ 15 inches Hg.