Connect the CPC to the pump inlet hosetail - the short
length of tubing supplied connected to the CPC is
connected to the pump hosetail. Ensure that the CPC is
connected correctly, the side of the CPC with the label
should be pointing away from the pump.
The low flow adapter can now be connected to the CPC
inlet hosetail. Set the low flow rate with the throttle valve on
the low flow adjuster, using a suitable flow calibrator (such
as a Defender primary calibrator).
Note: If carrying out high flow sampling subsequent to low
flow sampling with the same pump, ensure that the CPC
is disconnected.

Low Flow Sampling

CPC

Low flow
adaptor