Specifications, All models, 52mtx only – SKC Limited Sidekick MTX Pump Operating Instructions User Manual

Page 5

Advertising
background image

www.skcltd.com

224-SMTXM Issue F

Page 3

Specifications

All models

• Weight:

224-50MH - 465 g

224-51MTX & 224-52MTX - 575 g

• Enclosure rating: IP20
• Operating range: 5-3000 ml/min (5 - 500 requires adapter)
• Flow control: ±5% set point constant flow
• Battery pack: NiMH rechargeable, 2.0 Ah, 4.8 V
• Storage & operating temperature: -20°C to +40°C
• Operating humidity: 0 to 95% RH
• Compensation range:

Flow rate

(ml/min)

Back Pressure

(inches of water)

Typical sample media back pressure:

I.O.M. sampler with GFA filter at 3 l/min =

approximately 7 inches of water

I.O.M. sampler with 0.8µm MCE filter at 2 l/min =

approximately 15 inches of water

1000

5 - 25

2000

0 - 25

2500

0 - 20

3000

0 - 15

• Typical run time: 8 hours minimum at 3000 ml/min and 15 inches of water back pressure
• Low battery shutdown function

224-52MTX only:

Flow fault shutdown function in event of high back pressure or pinched hose

Timer:

Run time display 0 to 99,999 minutes

Accuracy ±0.05%

Timer freeze and run time retention on low battery and flow fault shutdown

79.5mm

57mm

145mm

Advertising