Preparing to take a sample, Setting the alarm – SKC Limited SPLIT2 User Manual

Page 14

Advertising
background image

14

Auto - Zeroing
Please Wait

24

Run
Review Data
Special Functions
Auto-Zero

The Split2 is now Auto - Zeroing. The time in seconds will appear
in the bottom right hand corner of the LCD indicating how long
for the Auto - Zero to be completed.

Once completed the SPLIT2 will return to the main menu

5. Setting the Alarm:

Go to the main menu screen. (As a starting point we recommend
the alarm is set at 80% of the Occupational Exposure Limit)

Run
Review Data
Special Functions
Auto-Zero

System Options
Date/Time
Set Alarm
Cancel

Scroll arrow to Set Alarm option and press the ENTER button

The ALARM is configured as concentration in Milligrams per cubic metre (Mg/m

3

)

To set all parameters on this screen the following procedure applies:

To increase the value of the selected digit... Press the UP button.
To decrease the value of the selected digit .... Press the DOWN button.

To confirm the digit and select the next field.... Press the ENTER button

When all the data is correct press ENTER The screen will now return to the Main Menu

Go to the Special Functions option by pressing either the down
button or the up button until the arrow is opposite Special
Functions. Press ENTER button

*

SET ALARM

*

Conc: 000.00 mg/m

3

Cursor

The following screen will appear with a cursor (line) under
the 0 character

Please NOTE: Alarm only works if the Alarm Continue
option is taken from the RUN menu. Please see section
on TAKING A SAMPLE page 17

PREPARING TO TAKE A SAMPLE