Consumables & spares temperature changes – SKC Limited SPLIT2 User Manual

Page 22

Advertising
background image

22

CONSUMABLES

IOM cassette, plastic

225-71A

Foam inserts for Respirable selection

225-772

Other selector foams not currently available ---
GFA 25 mm filters

225-58F

PARTS & ACCESSORIES

Sidekick air sampling pump

224-50

Sidekick air sampling pump with timer

224-52TX

(224-52TX is an intrinsically safe pump)
Charger for Sidekick

223-203C*

Tygon flexible tubing ( one metre)

225-13-4A

DC Lite calibrator

717-01K

Rotameters also available - see options in the latest catalogue
Calidaptor to connect DC Lite/rotameter

390-01

Replacement SPLIT2 Battery Pack

770-303

SPLIT2 Battery Charger

770-310C*

Replacement data cable

770-114

Replacement calibration standard

770-207

Replacement zero filter

770-112

Manual

770-300M

*C denotes a UK plug. For a European 2 pin plug use B

This section is IMPORTANT as it covers the procedure involved to ensure the reliability and accuracy of your unit is optimised despite
the often unavoidable changes in temperature.

Drastic changes in the environment temperature could affect the readings from the Split2. Temperature changes can occur, for example,
when transferring between indoor and outdoor monitoring, or if the unit has been stored overnight in a cold room or vehicle, before
being used in a warm room.

Follow this procedure:

1. Take the SPLIT2 to the area where you wish to sample
2. Run the unit for 15-20 minutes with the zeroing filter on and the perform an auto zero.
3. Remove zeroing filter.
4. Choose overwrite option. This will erase all the previous memory locations already stored and allow the electronics to

stabilize (to the sampling temperature). The auto zero will activate the temperature compensation.

It is recommended to auto zero the unit each time the user changes location, this resets the temperature compensation to
ambient

CONSUMABLES & SPARES

TEMPERATURE CHANGES