Procedure – Teledyne LeCroy HFP1500 User Manual

Page 31

Advertising
background image

Operator’s Manual

922252-00 Rev A

25

Procedure

A. Output Zero Voltage

Figure 5 Output Zero Voltage Test Setup

1. Connect one end of a BNC cable to the female BNC connector on the probe end of the

ProBus extender cable. Connect the precision 50 Ω terminator to the other end of the BNC
cable.

2. Connect the banana plugs of the Precision terminator to the input of the DMM. Make sure

that the plug corresponding to the BNC shield (marked "Ground") is connected to the LO or
COMMON input of the DMM. Refer to Figure 5 Output Zero Voltage Test Setup for setup
information.

3. Set the OFFSET on the oscilloscope to zero, as indicated by on-screen display.

4. Set the DMM to read DC Volt on the most sensitive range.

5. Record the voltage measured on the DMM to 10 μV resolution as ’Output Zero Voltage’ in

the Test record.

6. Check that the voltage indicated by the DMM is between ±800 μV.

7. Disconnect the DMM from the precision 50 Ω terminator. Leave the remaining setup in

place for the next step.