5 force-record shunt installation – Acumen INSTRUMENTS CORPORATION DATABRIDGE(TM) SDR2-USB User Manual

Page 27

Advertising
background image

SDR2-USB Configuration Guide rev 1.0

J4 Remote Control Port

25

Contact Function Direction

Wire

Color

1 V

DD

(3.3 VDC)

-

2

GND (0 VDC)

Black

3

input Blue

4 input

Grey

5

output Yellow

6 no

connect

-

7

output Orange

8 output

Green

9 V

DD

(3.3 VDC)

-

10

GND (0 VDC)

-

Table 5.2. J4 remote control port pinout.

J4 mating plug (shell): Hirose DF11-10DS-2C (Digi-Key part number
H2023-ND)
For use with DF11 crimp contacts e.g. DF11-2428SCA (Digi-Key part number
H2300-ND) or pre-crimped wires (Digi-Key part number H3BXG-101LL-CW
where LL=length in inches, C=color code, W=wire gauge code).

5.5 Force-Record Shunt Installation

The SDR2-USB can be forced into record mode any time power is applied
and USB storage is present. To use this mode, install a 2mm shunt between
J4 pins 4 (REC_TOGGLE) and pin 2 (GND).

A good choice of shunt is Norcomp part number 810-002-SP2L001 (Digi-
Key part number SP2-001E-ND).

REC_BUTTON

REC_TOGGLE

FAULT

DATA

RECORD

Advertising