it. When finished, turn the instrument off
and flip DIP switch #7 to “off.”

On the Model IAQ50, press and hold the
TEMP

or HUMIDITY key to enter the

calibration mode for that parameter. To
enter the calibration mode for the RTD,
make sure the RTD is plugged in. Press the
TEMP

key until “Prb” shows on the

display, then press and hold the TEMP
key.

The display will begin a countdown from
5 to 0. Release the key when 0 is displayed.
The top line of the display will show the
adjustment number (factory setting is 0.0)
and the bottom line will show the actual
reading value. Use

5 or 6 keys to change

the adjustment number and press ENTER
to accept. If necessary, repeat the above
sequence for the other parameters. When
finished, turn the instrument off and flip
DIP switch #7 to “off.”