Troubleshooting – Tracer TP-9363 PRO-Chek CG AM08045-4 User Manual

Page 9

Advertising
background image

9

Troubleshooting

WARNING
Substitution of components may impair intrinsic safety.

Except for the batteries and the sensor, the internal parts of the
PRO-Chek CG Combustible Gas Leak Detector are not user serviceable.
If you experience a problem with your PRO-Chek CG, follow the steps in
the troubleshooting guide below to determine how to remedy your problem.
If you cannot remedy the problem, contact the supplier for warranty
evaluation and instructions.

PROBLEM

POSSIBLE CAUSE

REMEDY

The PRO-Chek CG will

not power up.

Batteries are worn out.

Replace batteries.

Batteries are improperly

installed.

Refer to Figure 1 for

proper installation.

Poor sensitivity. Unit

powers up, but has a

difficult time detecting

combustible gas.

Sensitivity setting is

too low.

Increase sensitivity by

dragging finger upward on

touch pad.

Sensor is worn out and

needs to be replaced.

Replace sensor.

Alarm sounds

continuously, even after

a couple minutes of

warm-up.

The sensor is not installed

properly, is missing or is

worn out.

The sensor leads must

be straight and inserted

in the holes at the base

of the sensor socket.

Or, the orientation tab is

not properly aligned in

the probe tip. Refer to

Figure 2.

Sensor is worn out and

needs to be replaced.

Replace sensor.

Advertising