REV 2 PCI-600 USER’S MANUAL

6

2.0 OPERATING

PROCEDURES

2.1 Functional

Description

2.1.1. AC Current Source

AC test currents from 10A to 600A are set by turning the

[CURRENT CONTROL]

knob on the front panel. The test

current is measured and displayed on the LCD screen. The PCI-
600 output current ratings are shown in Table 3 below.

Table 3. Current Output vs. Time

Output @ 120 Vac

Output @ 240 Vac

Time

5.6 Vac @ 100 A

9.5 Vac @ 100 A

1 hour

5.3 Vac @ 200 A

9.4 Vac @ 200 A

5 minutes

4.9 Vac @ 300 A

9.0 Vac @ 300 A

2 minutes

4.6 Vac @ 400 A

8.2 Vac @ 400 A

1 minute

4.2 Vac @ 500 A

7.5 Vac @ 500 A

30 seconds

3.9 Vac @ 600 A

7.0 Vac @ 600 A

20 seconds

2.1.2. Current Output Control

The PCI-600 current source output is controlled by the

[↑]

and

[↓]

keys. Three control modes are available:

ON + TMR This mode turns on the PCI-600's current source

and timer. This initiates a test and is stopped by
using the "Timer Stop" inputs. The test results will
be displayed on the LCD. The test can be
terminated by pressing the

[↓]

key. The PCI-600's

built-in time/cycle counter is used to time events in
milliseconds and cycles. The elapsed time is
displayed on the LCD along with the test current
after a test is completed.

OFF

In this mode both the current source output and
timer are off.

MONT.

This mode turns on the current source
momentarily. To turn on the current source, press
and hold the

[↓]

key. The PCI-600 current output

can now be set by turning the

[CURRENT

CONTROL]

knob. Release the

[↓]

key to turn off

the current source.