Chg mo e, Cycle test – ADIC FastStor Mass Storage Device User Manual

Page 55

Advertising
background image

Troubleshooting and Diagnostics

39

continue this process until the volume is full. It is important to note

that while in Sequential-Access mode, the FastStor

robotics is

not logically connected to the SCSI bus and does not respond to SCSI commands.

After entering the dIAG MENU, press the MODE button until CHG MOdE appears on the LCD:

Chg mo e

2

3

4 5 6 7

Press the SELECT button to display the current mode.

Press the NEXT or PREVIOUS button to toggle the mode between SEQUENTIAL and RANDOM.

Press the SELECT button. CYCLE PWR will blink and appear on the LCD for approximately 10 seconds.

Press and hold the POWER button for two seconds to power down the FastStor.

Wait a few seconds, then press the POWER button again to power up the FastStor.

When in Sequential-Access mode, the

FastStor

replaces the LdR REAdY message with a SEQ REAdY message.

If you press MODE when the FastStor is in Sequential-Access mode, a STOP SEQ menu will appear on the LCD. Pressing
SELECT while in this menu will cause a SEQ ENdING message to replace the LdR REAdY message and Sequential mode will
finish after the currently loaded cartridge is unloaded. Pressing MODE after the cartridge is unloaded will cause FastStor to enter
the normal menu structure.

When the SEQ ENdING message is displayed, pressing MODE before the current cartridge is unloaded will cause a RESUME
SEQ
menu to be displayed. Pressing SELECT while in this menu will cause the previous STOP SEQ request to be canceled and
Sequential mode will continue with the current cartridge as if STOP SEQ had never been requested.

When the last cartridge in the volume has been filled, a SEQ dONE message will replace the SEQ REAdY message.

When the SEQ dONE message is displayed, pressing MODE will cause a START SEQ menu to be displayed. Pressing SELECT
while in this menu will cause Sequential mode to restart. Pressing MODE while in this menu will cause FastStor to enter the
normal menu structure.

CYCLE TEST

Notes

Before executing CYCLE TEST, verify that the drive is empty. If necessary, use Mode 1B – EJECT DRIVE, to remove any

cartridge from the drive.

Before executing CYCLE TEST, verify that all seven cartridge slots have tapes installed. If required, install additional

cartridges.

If AUTOCLEAN is enabled and a cleaning tape is in slot 7, CYCLE TEST will not use slot 7 during the test.

Advertising