After assessing the patient’s condition, sit the patient up and make a single cut down the back of the
patient’s clothing (see Figure 3-4), or undo any ties on the back of the gown in a hospital. The
posterior defibrillation/pacing pad may be placed on the patient's back at this time if the local
protocol calls for anterior-posterior placement. Use of standard pads in both the a-p or anterior-
anterior/apex-sternum placement is acceptable and will not affect the operation of the AutoPulse or
defibrillator.

5.

Slide the AutoPulse Platform into position behind the sitting patient and lay the patient down onto
the Platform. Placing the AutoPulse to the patient's side and "log rolling" him or her onto the
Platform is an acceptable alternative.

Figure 3-4

Cutting Patient Clothing and Positioning of AutoPulse