If the chest band is not completely unwound from the driveshaft follow this procedure

Figure 2-9

Positioning the LifeBand for Removal

1.

Once the cover plate is removed, take hold of the chest band on either side of the driveshaft and pull
both bands outward.

Figure 2-10 LifeBand in the Home Position: Ready for Removal

2.

The chest bands should now be fully extended. The seam securing the band clip should easily be
seen and the driveshaft is centered. The LifeBand may now be removed continuing with the
procedure in Section 2.1.2 on page 2-6. Do not remove the LifeBand if it is not in this position.

LifeBand Seam