The co-oximetry menu (spco/spmet installed) – ZOLL E Series Monitor Defibrillator Rev H SPO2 User Manual

Page 10

Advertising
background image

E Series - Pulse CO-Oximetry (SpO

2

, SpCO, and SpMet) Option Insert

8

9650-1211-01 Rev. H

The SpO2 Menu (SpCO/SpMet not installed)

When the SpO2 parameter is selected, the following
softkeys appear: Sens., Average and Return (See
“SpO

2

Settings softkey”, below).

Press the Return softkey to return to the main menu.

The CO-Oximetry menu
(SpCO/SpMet installed)

When the CO-Oxim. parameter is selected, the following
softkeys appear: SpO2 Settings, Enable Auto disp,
SpCO, SpMet, and Return.

Press the Return softkey to return to the main menu.

SpO2 Settings softkey

When you select the SpO2 Settings softkey, the
following softkeys appear: Sens., Average, and Return.

Sens. Softkey: The Sens. softkey allows you to select
either Normal or High sensitivity for SpO

2

monitoring.

The Normal sensitivity setting is the recommended
setting; select it for most patients.

The High sensitivity setting allows you to monitor
SpO

2

even under very low perfusion conditions. Such

conditions may include severe hypotension or shock.
With the High sensitivity setting, however, SpO

2

results are more easily contaminated by artifact. To
ensure accuracy of SpO

2

readings when using the

High sensitivity setting, observe the patient carefully
and continuously.

To select the sensitivity mode, press the Sens. softkey
to toggle between the modes, then press the Enter
softkey to select the highlighted mode.

Pressing the Return softkey returns to the SpO

2

Sensitivity/Average submenu without changing
sensitivity.

Average Softkey: The E Series provides 3 different time
periods over which SpO

2

values are averaged:

4 seconds, 8 seconds (default), and 16 seconds.

The averaging period is rarely changed from the
8 second default setting. For high risk patients with
rapidly changing SpO

2

conditions, use the 4 second

setting. Use the 16 second setting only when the
8 second setting (default) is inadequate due to extremely
high artifact conditions.

To select the averaging period (4, 8, or 16 seconds), first
press the Average softkey on the SpO

2

submenu. The

following softkeys appear: Average, Enter and Return.

Press the Average softkey again to scroll through the
different averaging periods, then press the Enter softkey
to select the highlighted averaging period.

Pressing the Return softkey to returns to the SpO2
Sensitivity/Average submenu without changing the
average.

Enable Auto disp softkey

This feature allows you to set the CO-oximetry data field
to alternate between SpO

2

, SpCO and SpMet readings,

if installed. SpO

2

will display for 20 seconds, followed by

SpCO for 10 seconds. SpO

2

will then display again for

20 seconds, followed by SpMet for 10 seconds (if
installed). This alternating pattern will repeat until Auto
Display is disabled. Press Enable Auto disp to enable
the feature; press Disable Auto disp to disable it.

If Auto Display is enabled, then the SpCO and SpMet
softkeys will not appear.

SpCO softkey

Pressing this softkey will instantly display the SpCO
reading in the CO-oximetry data field. It will display for 10
seconds, then revert to the SpO

2

reading. This softkey

will not appear if Auto Display is enabled.

Enter

Select

Return

ECG

ZDR Dest.

SpO2

NIBP

Enter

Select

Return

ECG

ZDR Dest.

CO-Oxim.

NIBP

Enable

Auto disp

SpO2

Settings

Return


SpCO

SpMet

Average

Sens.

Return


Enter

Sens.

Return

Normal

High

Enter

Average

Return

4 secs

8 secs

16 secs