Default settings, Co-oximeter accessories – ZOLL E Series Monitor Defibrillator Rev H SPO2 User Manual

Page 12

Advertising
background image

E Series - Pulse CO-Oximetry (SpO

2

, SpCO, and SpMet) Option Insert

10

9650-1211-01

Rev.

H

Default Settings

When the pulse CO-oximeter is turned on, the following default settings are automatically selected and remain in
operation until changed.

Note: Only the default settings are available in AED Semi-Automatic operation. Default alarm limit setting may be

adjusted in System Configuration mode. See the E Series Configuration Guide for more information.

CO-Oximeter Accessories

The following table describes each of the SpO

2

, SpCO, and SpMet accessories.

Note: LNCS single use and reusable sensors are capable of measuring SpO

2

only. Rainbow single use sensors are

capable of measuring SpO

2

, SpCO, and SpMet.

Parameter

Default Setting

Range

Averaging Mode

8 seconds (medium)

4 seconds (short)

8 seconds (medium)

16 seconds (long)

Sensitivity

Normal

Normal or High

High SpO

2

Alarm Limit

100%

72% to 100%

Low SpO

2

Alarm Limit

85%

70% to 98%

High SpCO Alarm Limit

10%

2% to 100%

Low SpCO Alarm Limit

0%

0% to 99%

High SpMet Alarm Limit

3.0%

1% to 100%

Low SpMet Alarm Limit

0.0%

0% to 99%

High Heart Rate Alarm Limit

150 bpm (beats/minute)

60 to 280 bpm (beats/minute) - monitoring via
ECG

60 to 235 bpm (beats/minute) - monitoring via
Pulse Oximetry

Low Heart Rate Alarm Limit

30 bpm (beats/minute)

20 to 100 bpm (beats/minute)

Item

Description

REF

LNCS Adtx

Single use sensor for patients > 30 kg

8000-0320

LNCS Pdtx

Single use sensor for Pediatrics and Slender Adults 10 - 50 kg

8000-0321

LNCS Inf-3

Single use sensor for Infants 3 - 20 kg

8000-0322

LNCS Neo-3

Single use sensor for Neonates < 3 kg

8000-0323

LNCS NeoPt-3

Single use sensor for Neonates < 1 kg (Pre-term)

8000-0324

LNCS DCI

Reusable sensor for Adults and Pediatrics > 30 kg

8000-0294

LNCS DCIP

Reusable sensor for Pediatrics 10 - 50 kg

8000-0295

Red LNC-04

4' Reusable Patient Cable - connects to LNCS single use and reusable
sensors listed above for monitoring SpO

2

.

8000-0330

Red LNC-10

10' Reusable Patient Cable - connects to LNCS single use and reusable
sensors listed above for monitoring SpO2.

8000-0331