Ensuring accurate monitoring – ZOLL E Series Monitor Defibrillator Rev H SPO2 User Manual

Page 7

Advertising
background image

Ensuring Accurate Monitoring

9650-1211-01 Rev. H

5

cable’s male connector all the way into the patient
cable connector (see Figure 5).

Figure 5

5. Lower the clear plastic protective cover over the

connection to secure it (see Figure 6).

Figure 6

6. Connect the patient cable to the CO-Oximetry

connector on the back of the E Series unit as shown
in Figure 2 on page 4.

Applying a Disposable Sensor

You can use a disposable LNCS sensor for SpO

2

monitoring or a disposable Rainbow sensor to monitor
SpO

2

, SpCO, and SpMet. Do not wrap the adhesive too

tightly as this can cause venous pulsations that could
lead to inaccurate saturation measurements.

You can reapply a disposable sensor to

the same patient

if the emitter and detector windows are clear and the
adhesive still adheres to the skin.

You can partially rejuvenate the adhesive by wiping it
with an alcohol wipe and allowing the sensor to
thoroughly air dry prior to replacement on the patient.

Note: LNCS Adult disposable sensors are not intended

for use across a child's hand or foot. For
instructions on proper application of neonatal
sensors, refer to the Directions for Use included
with each LNCS sensor.

1. Open the pouch, and remove the sensor.

2. Holding the sensor with the printed side downward,

bend the sensor backward and remove backing
material from the sensor.

3. Orient the sensor so that the digit can be attached to

the detector side of the sensor first (see Figure 7).

Figure 7

4. Press the detector onto the fleshy part of the finger

near the tip of the finger. To ensure accurate data,
you must have complete coverage of the detector
window.

5. With the emitter positioned over the fingernail, secure

the wings around the finger (see Figure 8).

Figure 8

When positioned properly, the:

emitter and detector are vertically aligned

digit completely covers the detector window

connector tab is located on the top side of the
finger

6. Lift the clear plastic protective cover from the female

end of the patient cable, then plug the sensor cable’s
male connector all the way into the patient cable
connector (see Figure 5).

7. Lower the clear plastic protective cover over the

connection to secure it (see Figure 6).

8. Connect the patient cable to the CO-Oximetry

connector on the back of the E Series unit as shown
in Figure 2 on page 4.

Note: For the Rainbow patient cable and Rainbow

disposable sensor, plug the sensor cable’s male
connector all the way into the patient cable
connector until it clicks.

Ensuring Accurate Monitoring

The following points aid in ensuring oximetry monitoring
success:

Choose a site that is well perfused and allows for
proper alignment of the light emitter and detector.

Select a site that has unrestricted blood flow.

ZM 040022A

Sensor Cable Connector

Patient Cable
Connector

Clear Protective
Cover

ZM 040024A

Detector Window