Preparing to Take Measurements

ZOLL M Series Operator’s Guide - Option Insert

7

WARNING

Do not start NIBP measurements unless you are sure that the cuff inflation and alarm settings are

appropriate for the patient. Incorrect settings can result in patient injury or inaccurate measurements.

Displaying the NIBP Menu

Unless it is an emergency situation where quick reaction is essential, you should always check that the
cuff inflation and alarm settings are set properly before taking a measurement. You can access all
NIBP features through the NIBP menu (see Figure 1 on page 3). You can display the NIBP menu
without leaving Monitor, Defib or Pacer mode. AEDs must be in manual override mode to display the
NIBP menu.

Displaying the
NIBP Menu

To display the NIBP menu:
1. Press the Param softkey. If the Param softkey label is not displayed, press the Return softkey

until “Param” displays.
The M Series unit displays the Parameter menu (see the following figure):
Figure 3 Parameter Menu

2. Press the Select softkey until “NIBP” highlights.
3. Press the Enter softkey.

The M Series unit displays the NIBP menu:
Figure 4 NIBP Menu

On M Series CCT units, the following NIBP menu appears:
Figure 5 NIBP Menu (CCT only)

Table 3 Accessing NIBP Features

Task

Action

Taking a single measurement.

Press the NIBP button.

Taking STAT measurements.

Press and hold the NIBP button for two seconds.

Taking automatic measurements.

See "Displaying the NIBP Menu" on page 7.

Aborting measurement in progress.

Press the NIBP button.

Changing NIBP settings.

See "Displaying the NIBP Menu" on page 7.

Enter

Select

Return

SpO2

EtCO2

NIBP

NIBP

Auto

NIBP

Stat

Return

Cuff

Inflation

Auto

Interval

NIBP

Auto

NIBP

Stat

Return

NIBP

PT. Type Settings