Setting alarms, Setting alarm limits – ZOLL M Series Defibrillator Rev G NIBP User Manual

Page 17

Advertising
background image

Preparing to Take Measurements

ZOLL M Series Operator’s Guide - Option Insert

11

Setting Alarms

Before taking a measurement, ensure that all the NIBP alarms settings are appropriate for the
situation. See the M Series Operator’s Guide for instructions on setting alarms. When an NIBP or
heart rate alarm is triggered, the unit automatically initiates a single blood pressure measurement. See
the M Series Configuration Guide for instructions on how to enable/disable this automatic
measurement.
The M Series NIBP option includes blood pressure alarms for the following measurements:

* Neonatal and pediatric NIBP modes available on M Series CCT units only.

When the M Series unit is turned on, all alarm functions are disabled and the alarm limits are set to
their default values. Any changes to these settings remain in effect until either the settings are changed
or 10 seconds after the M Series unit is turned off. On M Series CCT units, you can configure the
default alarm limits at power-up according to patient type; see the M Series Configuration Guide for
more information.

Setting Alarm
Limits

You can change all the alarm limits using standard M Series procedures. See the M Series Operator’s
Guide
for instructions on setting alarm limits.
You can set the alarm limits using the Auto function, or you can choose your own limits (see Table
4 on page 11
). The Auto function sets the high limit to 30 mmHg (4.0 kPa) above and the low limit to
20 mmHg (2.7 kPa) below the last measurement.

Table 4 Alarm Limit Settings and Defaults

Patient Type

Alarm Parameter

Default Setting

Range (increments of 5)

Adult

Systolic

High

160 mmHg (21.3 kPa)

80-260 mmHg (10.7-34.7 kPa)

Low

90 mmHg (12.0 kPa)

40-140 mmHg (5.3-18.7 kPa)

Diastolic

High

110 mmHg (14.7 kPa)

50-200 mmHg (6.7-26.7 kPa)

Low

50 mmHg (6.7 kPa)

25-100 mmHg (3.3-13.3 kPa)

Mean

High

130 mmHg (17.3 kPa)

60-220 mmHg (8.0-29.3 kPa)

Low

60 mmHg (8.0 kPa)

30-120 mmHg (4.0-16.0 kPa)

Pediatric*

Systolic

High

145 mmHg (19.3 kPa)

80-160 mmHg (10.7-21.3 kPa)

Low

75 mmHg (10.0 kPa)

35-140 mmHg (4.7-18.7 kPa)

Diastolic

High

100 mmHg (13.3 kPa)

50-130 mmHg (6.7-17.3 kPa)

Low

35 mmHg (4.7 kPa)

20-100 mmHg (2.7-13.3 kPa)

Mean

High

110 mmHg (14.7 kPa)

60-140 mmHg (8.0-18.7 kPa)

Low

50 mmHg (6.7 kPa)

20-120 mmHg (2.7-16.0 kPa)

Neonate*

Systolic

High

100 mmHg (13.3 kPa)

60-130 mmHg (8.0-17.3 kPa)

Low

50 mmHg (6.7 kPa)

25-120 mmHg (3.3-16.0 kPa)

Diastolic

High

70 mmHg (9.3 kPa)

30-105 mmHg (4.0-14.0 kPa)

Low

30 mmHg (4.0 kPa)

20-100 mmHg (2.7-13.3 kPa)

Mean

High

80 mmHg (10.7 kPa)

35-110 mmHg (4.7-14.7 kPa)

Low

35 mmHg (4.7 kPa)

20-105 mmHg (2.7-14.0 kPa)

Advertising