ZOLL M Series Operator’s Guide - Option Insert

* Neonatal and pediatric NIBP modes available on M Series CCT units only.

Table 11 Battery Operating Time Specifications

For a new, fully charged battery
pack at 20°C with NIBP, EtCO

2

and

SpO

2

Options

35 defibrillator discharges at maximum energy (360J), or 1.25 hours
minimum of continuous ECG monitoring and blood pressure
measurements once every 5 minutes, or 0.75 hour of continuous ECG
monitoring/pacing at 60 mA, 70 beats/min.

Table 12 Environmental Specifications

Operating Temperature

0

° to 50° C

Storage and Shipping Temperature

-20

° to 70° C

Electromagnetic Immunity

AAMI DF-2: 1996, EN 61000-4-3: 1996, 20 V/m

Humidity

5 to 95% relative humidity, non-condensing

Operating Pressure

594 to 1060 mBar

Table 13 Alarm Limits Specifications

Systolic

Adult

High 80 – 260 mmHg;
Low 40 – 140 mmHg;

High 10.7 – 34.7 kPa
Low 5.3 – 18.7 kPa

Pediatric*

High 80 – 160 mmHg;
Low 35 – 140 mmHg;

High 10.7 – 21.3 kPa
Low 4.7 – 18.7 kPa

Neonatal*

High 60 – 130 mmHg;
Low 25 – 120 mmHg;

High 8.0 – 17.3 kPa
Low 3.3 – 16.0 kPa

Diastolic

Adult

High 50 – 200 mmHg;
Low 25 – 100 mmHg;

High 6.7 – 26.7 kPa
Low 3.3 – 13.3 kPa

Pediatric*

High 50 – 130 mmHg;
Low 20 – 100 mmHg;

High 6.7 – 17.3 kPa
Low 2.7 – 13.3 kPa

Neonatal*

High 30 – 105 mmHg;
Low 20 – 100 mmHg;

High 4.0 – 14.0 kPa
Low 2.7 – 13.3 kPa

Mean

Adult

High 60 – 220 mmHg;
Low 30 – 120 mmHg;

High 8.0 – 29.3 kPa
Low 4.0 – 16.0 kPa

Pediatric*

High 60 – 140 mmHg;
Low 20 – 120 mmHg;

High 8.0 – 18.7 kPa
Low 2.7 – 16.0 kPa

Neonatal*

High 35 –110 mmHg;
Low 20 – 105 mmHg;

High 4.7 – 14.7 kPa
Low 2.7 – 14.0 kPa