Transmitting trend history or pc card data, Rs-232 e – ZOLL M Series Defibrillator Rev E Bluetooth User Manual

Page 10

Advertising
background image

B

LUETOOTH

/RS-232 E

NHANCED

D

ATA

C

OMMUNICATION

6

www.zoll.com

9650-0307-01 Rev. E

Transmitting Trend History or PC Card Data

You can transmit all trend history data stored in memory or all data stored on a PC data card

to a remote handheld device from System Utilities mode.
For detailed instructions on uploading data from a PC data card through an RS-232 serial link,

see “Appendix B” of the M Series Operator’s Guide.

Procedure
Follow these steps to initiate trend history or PC card data transmission via the Bluetooth

module to the remote handheld device.

Step Action

Result

1.

Turn the selector switch to OFF.

After 10 seconds, press and hold the
left-most softkey and turn the
selector switch to MONITOR (ON for
AED units).

The M Series unit powers on in System Utilities mode.

2.

Press the Upload Trend or Upload
Card
softkey.

The M Series unit displays the Upload screen.

3.

Press the Send softkey.

The Bluetooth module’s green power LED illuminates.
The M Series unit establishes a connection to the
remote handheld device, and begins data transmission.
The unit displays a progress bar on the Upload screen.

Erase

Cards

Print

Config

Clock

Synch

Upload

Card

Upload

Trend

Press a Key or

Power Off to Exit

System Utilities

Return

Send

Approx. Upload Time

00:00:02

Upload

Uploading

Upload

Power Off to Abort

20%

Advertising