RS-232 serial port

On M Series units not equipped with a Bluetooth module, the RS-232 port on the back of the

unit provides serial communications capability to the remote handheld device or PC.

Caution The RS-232 serial port/ECG output connector does not function on an M Series unit

that is equipped with a Bluetooth module; you must use the RS-232/ECG output jack

on the Bluetooth module for serial data communications/ECG output.

Figure 2: M Series unit with RS-232 serial port

For data transmission through the RS-232 serial port, you must use a ZOLL M Series RS-232

data transfer cable (ZOLL part number 8000-0605-01) and a serial data cable that is

compatible with the remote handheld device.
The M Series data transfer cable terminates in a female DB9 connector; the serial data cable

that connects to the remote handheld device must terminate in a male DB9 connector. You

may also need a null modem adapter to connect the two cables; see the RescueNet Code

Review Utilities documentation for more information.

Communication Settings

You must configure the same communication settings on both the M Series unit and the

remote handheld device for proper data transmission. The remote handheld device may also

require a PIN code to connect to the M Series unit: in this event, use the default value of 0.

The remote handheld device must be running ZOLL Data Relay software to access any

information transmitted from the M Series unit. See the ZOLL RescueNet Code Review

Utilities documentation for installation and operating instructions.
Note: If you are transmitting data through the Bluetooth module and then change to serial

data transmission, you must cycle the power on the Bluetooth module (by pressing

the Data Relay On and Data Relay Off softkeys) before connecting the data

transfer cable. Failure to do so will result in a SERIAL FAULT error on the first

attempt at serial data transmission.

23 

Parameter

Value

Baud rate (bits per second)

115 200 bps
38 400 bps (default)
9600 bps

Data bits

8

Parity

None

Stop bits

1

Flow control

None