Set time, Changing paper, Setting time and date – ZOLL M Series Defibrillator Rev YH User Manual

Page 66

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

10-4

Changing Paper

Press down and pull the paper tray drawer where the
“RELEASE” label is located. The drawer slides open.

Check for adequate paper supply. If paper supply is low,
remove paper in the tray. Place a new pad of thermal
paper in the drawer with the paper coming off the top of
the pad and the grid facing up.

Pull enough paper off the pad so that the paper extends
out of the strip recorder when the paper compartment
drawer is closed.

Close paper compartment drawer by pushing the drawer
in and pressing down lightly where the “RELEASE” label
is located until the drawer is flush with the front of the
device.

Setting Time and Date

Check the time and date on the recorder annotation. If it
is not correct, set as follows:

Turn the SELECTOR SWITCH to OFF. The device must
remain off for at least 10 seconds before entering the
“Set Time” screen for setting the time manually as
described in the subsection “Manual Method” below, or
in Utilities Mode for setting the time automatically. See
the subsection “Automated Method” below for
instructions on setting the time automatically.

Manual Method

1. Press and hold the right-most softkey on the unit

while turning the SELECTOR SWITCH to the
MONITOR or ON position. When the “Set Time”
screen appears on the monitor, release the softkey.

2. The month field will be highlighted. Press the Inc

Value or Dec Value softkeys to select the appropriate
month. Pressing the Next Field softkey will set the
selected month and move the highlight to the next
field (day).

3. Repeat above steps to set the correct day, year,

hours and minutes field.

Note: The last field does not automatically scroll (wrap)

to the beginning. You must press the Prev Field
softkey to enter the values for the last field.

If you need to make corrections, press the Prev
Field
softkey to move the highlight to the field
previously entered.

4. Press the Enter and Return softkey to set all values

and return to normal monitoring mode.

Note: Repeated display of the “SET CLOCK” or

“CLOCK FAULT” message may indicate that the
internal battery powering the unit’s Real Time
Clock is depleted. See the Troubleshooting
Guides
in Section B of this manual for more
information.

Automated Method

Note: This method requires a modem connection. In

addition, the correct time zone must be set in
System Configuration mode for the updated date
and time to be correct.

1. Press and hold the left-most softkey on the unit while

turning the SELECTOR SWITCH to the MONITOR
or ON
position. When the “System Utilities” screen
appears on the monitor, release the softkey.

2. Press the Clock Sync softkey. A setting screen

appears, allowing the user to choose a NIST
(National Institute for Science and Technology)
location and a prefix for the phone number of the
selected NIST location, as required. For example, if
the NIST location is outside of the local calling area,
users within the continental United States would
enter a “1” as the dial prefix. Other users would enter
a dial prefix as required for placing calls in the
continental United States.

3. Press the Dial softkey. The word “Initializing” appears

briefly, followed by the “Clock Synchronization”
screen, displaying the user configurable NIST phone
number with the appropriate prefix. The word
“Dialing” appears underneath, along with a seconds
counter, as the unit connects to the NIST site.

After receiving the atomic clock information from the
NIST site, the unit then displays updated date and time
information, unless one of the following errors occurs:

Set Time

Prev

Field

Enter

and

Return

12, 01

16:37

Inc

Value

Dec

Value

Next

Field

May

Error Message

Description/Corrective Action

MODEM REQUIRED

The unit determined that there is no
modem card installed. Install a
supported modem card and retry.

MODEM INIT ERROR The modem could not be

initialized. Replace the modem card
and retry.

NO DIAL TONE

The unit could not detect a dial
tone. Check the connection
between the modem and the phone
jack, or try a different phone line.

BUSY

The unit detected a busy signal
from the selected NIST location.
Retry.

NO ANSWER

The unit received no answer from
the selected NIST location. Retry or
select another NIST location.