34. “LOW BATTERY” message appears on monitor
when unit is plugged into AC mains.
•
Replace battery pack with a fully charged battery pack.
•
Unplug device from AC mains and plug device back into AC
mains.
•
Verify AC mains is working properly.
35. None of the CHARGER ON indicators are
illuminated when the device is plugged into AC
mains.
•
Unplug device from AC mains and plug device back into AC
mains.
•
Verify AC mains is working properly.