AC Charger

27. “CHECK PADS” message.

Verify proper Multi-Function Cable / MFE Pad connection by
disconnecting and reconnecting the Mult-Function cable and
MFE Pads.

Ensure proper contact of Multi-Function Pads and that the
patient does not have excessive hair beneath the electrodes.

If message persists, disconnect Multi-Function cable from
MFE Pads and plug cable into test connector. “CHECK
PADS” should change to “DEFIB PAD SHORT” (Manual
Mode Only).

If test fails, try using paddles to defibrillate.

28. “USE PADS” message.

ECG Analysis will only operate when MFE pads are attached
to the patient.

Disconnect paddle and connect MFE Pads for use in Semi-
automatic defibrillation.

Activate manual mode to use Paddles.

29. “NOISY ECG”

“RETRY ANALYSIS” message.

Check for proper application and adhesion of Multi-Function
Electrodes.

Check to make sure that nobody is touching the patient and
that the patient is motionless.

30. “ECG TOO LARGE”

“RETRY ANALYSIS” message.

Press ANALYZE button again to begin analysis.

31. No “TEST OK” message when performing a

defibrillator self-test.

Check to make sure unit is set to 30 joules.

If testing with Multi-Function Cable, make sure that cable is
firmly inserted into test connector.

If testing with paddles, make sure to press the paddles firmly
against the sides of the unit while discharging.

32. “DEFIB MAINT. REQUIRED” message.

Contact ZOLL Technical Service Department.

Symptom

Recommended Action

33. The green and orange-yellow CHARGER ON

indicators are alternately illuminating.

Verify battery is installed.

Turn unit ON to identify the fault condition.

Replace battery pack with a fully charged battery pack.

If problem persists, replace battery pack, unplug device from
AC mains and plug device back into AC mains.

Symptom

Recommended Action