IM 01E20C02-01E

4-7

4. WIRING

4.4.6

Wiring the Remote Flowtube
with the AXFA14 Converter

WARNING

Before wiring, be sure that the power supply for
AXFA14 converter has been turned off to
prevent an electrical shock.

(1) Connection with the Remote Flowtube

(General-Purpose Use, Submersible Type,
Sanitary Type, Size 2.5 to 400 mm (0.1 to
16 in.))

Connect wiring as shown in the figure below.

AXFC dedicated
signal cable

EX2

EX1

A

B

C

Excitation cable

EX1

EX2

C

SA

A

B

SB

F0413.EPS

Converter

Remote

flowtube

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

AXFA14 Converter

Remote flowtube

Figure 4.4.5

Wiring Diagram

(2) Connection with the Remote Flowtube

(Explosion proof Type, Size 2.5 to 400 mm
(0.1 to 16 in.))

In case of explosion proof type for ATEX, FM, CSA,
IECEx and TIIS certification, connect wiring as shown
in the figure below.

In case of the explosion proof type, the protective
grounding

of remote flowtube must be connected to

a suitable IS grounding system. In that case,
(functional grounding terminal) need not be connected.

AXFC dedicated
signal cable

EX2

EX1

A

B

C

Excitation cable

EX1

EX2

C

SA

A

B

SB

F0414.EPS

Converter

Remote

flowtube

SA

A

B

SB

C

EX1

EX2

Taping*

A

B

Taping*

C

EX1

EX2

* Individually tape and insulate the

shields corresponding to SA and
SB on the remote flowtube side.

AXFA14 Converter

Remote flowtube

Figure 4.4.6

Wiring Diagram

4.4.7

Connecting to External Instru-
ments

WARNING

Before wiring with external instruments, be sure
to turn off the power supply for AXFA14
converter and any external instruments.

Connect the AXFA14 terminal to external instruments,
giving attention to the following points.

For F

OUNDATION

Fieldbus protocol, please refer to IM

01E20F02-01E.

For PROFIBUS PA protocol, please refer to IM
01E20F12-01E.

4 to 20 mA DC Current Output

Resistive Load Max. 750

Ω.

Receiver

Instrument

AXFA14

l+

l-

F0415.EPS

Communication Resistive load

BRAIN

HART

250 to 450

250 to 600

Figure 4.4.7

4 to 20 mA DC Output Connection