8 installing the cover – Yokogawa AXFA14G/C User Manual

Page 15

Advertising
background image

IM 01E20C02-01E

4-9

4. WIRING

This output cannot switch an AC load. To switch
an AC load, an intermediate relay must be
inserted as shown in Figure 4.4.11.
*The alarm output operates from open (normal)
to closed (alarm occurrence) in the default value
(as setup upon plant shipment). Changes can
be made via the parameter settings.

F0419.EPS

Load

Protective diode

External power supply
30V DC, 0.2A. max

AXFA14

AXFA14

This connection is not possible.

DO+ (or DIO+)

DO- (or DIO-)

DO+ (or DIO+)

DO- (or DIO-)

Electromagnetic

valve

AC power supply

Relay

Figure 4.4.11 Status Output/Alarm Output Connection

NOTE

For status and alarm outputs from the DO or
DIO terminals, parameters must be set. Refer to
“Chapter 6: Parameter Description” in this
manual.

4.4.8

Installing the Cover

Install the cover to the flowmeter by turning it in the
direction of the arrow as shown below. Tighten cover
locking screw 2 counterclockwise using a hexagonal
wrench (nominal size 3 mm) to lock the cover.

F0420.EPS

Cover locking
screws

Figure 4.4.12 Installing the Terminal Box Cover