General Gas

Gases:

Handles most gases for which the critical
temperature, pressure and SG are known.

Compressibility:

Calculated using Redlich-Kwong equation.

Display:

Corrected Volume (m

3

or ft

3

).

Mass (kg or lbs).

Temperature Range:

-273°C (-450°F) to 800°C (1472°F).
(RTD has a more limited range.)

Pressure Range:

0 kPa abs (0 psia) to 100,000 kPa (10,000
psia).

Natural Gas

Calculations:

Uses NX-19 equation to calculate

supercompressibility Fpv.

Displays:

Corrected Volume (m

3

or ft

3

).

Mass (kg or lbs).

Temperature Range:

-40°C (-40°F) to 115°C (240°F).

Pressure Range:

101.325 kPa (14.69 psia) to 34,380 kPa (4985

psia).

SG Range:

0.554 to 1.000.

Carbon Dioxide:

0 to 15% mole.

Nitrogen:

0 to 15% mole.

Steam

Displays;

Mass (kg or lbs)

Energy (MJ or BTU x 1000).

Calculations:

Uses 1967 IFC Formulation equations to
calculate specific weight and enthalpy of
steam.

Steam Type:

Saturated and Superheated.

Temperature Range:

20°C (68°F) to 800°C (1472°F).
(RTD has a more limited range.)

Pressure Range:

1 kPa (abs) (1 psia) to 100,000 kPa (10,000
psia).