<3. Functions>

3-3

IM 11B06E01-01E

Read and write is individually performed, so that it is possible to offer one read request and one

write request at the same time.

Coil (00001 to 09999)

Data

Address

Description

Time setting

request

00001

For all GC8000 or GC1000 MarkII connected to ASGW

Time data should be set at the holding registers 40001 to 40004 before

request.

Value turns to “0” after writing.

To set time to all analyzers or PCAS, #TSET must be specified as 255..

Individual time

setting request

00nnn

Time is set for selected GC8000 or GC1000 MarkII.

Time data should be set at the holding registers 40001 to 40004 before

request.
“nnn” is 100 + StnID *

“nnn” (= 101 to 354)

Value turns to “0” after writing.

The setting for ASIU is invalid.

Mapping area

01001 to 09000 Setting the coil or input relay address of analyzers.

ASGW only reads the input relay on analyzers and only writes the coil on

analyzers.

Available is the following coils

GC8000

Refer to “5.1 Coils (command contacts)” in “GS 11B08B02-01E

GC8000 Modbus Communication”. However, “00004: Time Setting

Request” is not available.

GC1000 MarkII

Refer to “5.1 Coil (command contact)” in “GS 11B03G02-02E GC1000

MarkII Modbus Communication”. However, “00004: Time Setting

Request” is not available.

ASIU:

00101 to 00116: On command for DO01 to DO16

00201 to 00216: Off command for DO01 to DO16

Setting the other address is ignored.

This value turn to “0” after writing.

*:

StnID is the identification number of an analyzer. StnID is related to an IP address in the mapping file (see 4.2 Data mapping).

Input relay (10001 to 19999)

Data

Address

Description

In operation

10xxx

The value indicates “1” when the analyzer and PCAS is communicating.

“xxx” is StnID* of the analyzer.

“xxx” (=001 to 254)

Communication

error

10yyy

The value indicates “1” when there is an error in communication with ana-

lyzers and PCAS. Input relay or inout register needs to be defined.

“yyy” is 300 + Stn ID

“yyy” (= 301 to 554)

Write error

10zzz

The value indicates “1” when there is an error in writing data to analyzers

and PCAS; it returns to “0” when the next writing successfully finishes.

“zzz” is 600 + Stn ID

“zzz” (= 601 to 854)

Mapping area

11001 to 19000 Setting the input relay address of analyzers and PCAS.

ASGW reads the addressed data from analyzers and PCAS.

*:

StnID is the identification number of an analyzer. StnID is related to an IP address in the mapping file (see 4.2 Data mapping).

4th Edition : Nov. 22, 2011-00