Arcnet, Arcnet -1 – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 39

Advertising
background image

<7. ARCNET>

7-1

IM 11B06E01-01E

7. ARCNET

The New ARCNET-based bus system enables connection between an ARCNET system

and an Ethernet system. This realizes addition of Ethernet analyzers to existing ARCNET

systems.
A typical network configuration of analyzer bus is shown in Figure 7.1.
The network consists of analyzers (Gas chromatograph etc.) and "Analyzer bus sys-

tems for ARCNET". This "Analyzer bus systems for ARCNET" consist of “Gateway

unit”,”analyzer server” and “ARCNET / Ethernet Converter”.
Each software (PCAS, ASET) for “Analyzer server” and “Analyzer Server Engineering

Terminal” should be used ARCNET edition “/ARC”.
The communication to the analyzer connected to Ethernet is also enabled.

• "Analyzer bus systems for ARCNET" should be used so that DCS communicates

with analyzer connected to ARCNET.

• Redundancy for ARCNET is enabled. Redundancy for Ethernet can be done by using

double "Analyzer bus systems for ARCNET".

• One "Analyzer bus systems for ARCNET" can communicate with up to 30 sets of

analyzers connected to ARCNET. In case of communicating with more than 31 sets,

plural number of "Analyzer bus systems for ARCNET" should be used.

F0701.ai

DCS

DI/O

RS-232C

LAN (Ethernet)

Analyzer bus system

for ARCNET

ARCNET

PC

[PCAS/ARC]

PC

[ASET/ARC]

STARDOM FCJ

[ASGW]

ASIU

GCIU

GC1000 MarkII

GC1000

GC1000

[ ]:Software

ARCNET/Ethernet

Converter

GC1000 MarkII

GCIU

Analyzer

pH analyzer,

Oxygen analyzer, etc.

Network monitoring,

Automatic data storage

Gateway

Operation and display of

Gas Chromatograph

Figure 7.1

Typical System configuration

4th Edition : Nov. 22, 2011-00