background image

<5. Diagnostic Window>

5-4

IM 11B06E01-01E

5.3

Checking the analyzer communication

status and displaying the mapping data

The communication status window is displayed when [Modbus I/F] - [comm.status] is clicked.

UPDATE:

The data is updated.

ID:

StnID (define in the mapping file)

TYPE:

GC8000/GC1000/PCAS/ASIU (defined in the mapping file)

IP Addr:

IP address (define in the mapping file)

TASK:

The status is “OK” when the communication task is in operation.

COMM:

The status is “NG” when the communication error occurs.

WRITE:

The status is “NG” when it fails the writing, but is “OK” when it successes.

READ OK:

Number of success in reading

READ NG:

Number of failure in reading

WRITE OK:

Number of success in writing

WRITE NG:

Number of failure in writing

The mapping data window is displayed when [Modbus I/F] - [mapped data] is clicked.