2. Specifications

2.1 Standard Specifications

Measurement object

: Oxygen concentrations in inert gases containing no flammable gas, silica, corrosive gas, or liquid

(including water vapor).

Measurement system

: Zirconia system

Sampling method: Pump, aspirator, or no suction device.

Pump and aspirator suction flow rate
: Approx. 1.0 l/min.

Aspirator suction conditions

: Air or N

2

, supply pressure 65 to 100 kPaG, total discharge flow 10 l/min max. (when gas inlet and

outlet are at ambient atmospheric pressure).

Sample gas conditions

Flow rate

: 200 ± 25 ml/min (only applies to sensor).

Temperature

: 0-50°C (non-condensing).

Humidity

: Non-condensing.

Pressure

: 0-300 PaG

Measurement range

: 0-10 ppm O

2

to 0-100 vol% O

2

Resolution

: 0.01 ppm O

2

.

Display

: 4 digit LED.

Main display: O

2

concentration (auto switching).

Sub display: Parameter or alarm/error number

Unit

: %, ppm.

Output range

Auto

: 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, 0-100% (default)

Other : 0-0 ppm, 0-00 ppm, 0-000 ppm, 0-%, 0-0% can be set,

is an integer from 2 to 9.

Fixed

: Set to 0-10 ppm, 0-100 ppm, 0-1000 ppm, 0-1%, 0-10%, or 0-100%.

Partial

: Lower value or upper value of range can be set.

Note : Span (upper value-lower value) is 20% FS or more of above fixed range.

Example: 200-400 ppm when fixed range is 0-1000 ppm, 20-40 ppm when fixed range is 0-

100 ppm.

Analog output

: 2 outputs

Primary

: 4 to 20 mA DC (maximum load resistance: 550Ω )

Secondary

: Select from 0-1, 0-5, 0-10 V DC ( load resistance: 10 kΩ or greater)