3.2 How to Install

3.2.1 Installing Desktop Type

Place and operate the OX400 on a level surface as shown in Figure 3.1.
(1) Provide a distance of 100 mm or more behind the OX400 in order to not block the outlet of the

cooling fan on the rear panel.

2) The air inlet of the cooling fan is located on the bottom panel of the OX400. Be sure to provide

a distance from the installation surface (desktop) larger than that of the height of the legs of the

OX400.

20

Hot air is discharged from the air outlet on the rear panel of the OX400.

Maintain at least 100 mm of free space around the OX400 to ensure

adequate ventilation.

Notes on installation

Cooling fan

Unit: mm

Do not block the air inlet.

Distance of

20 or more

Air is discharged

from the cooling

fan

Distance of 100 or more

Figure 3.1

Notes on Installing Desktop Type