4. Piping and Wiring>

4-5

IM 11M10B01-01E

Table 4.3

Measurement Flow Path Switching Output by Multi-selector

MS1-MSCOM

MS2-MSCOM

MS3-MSCOM

MS OFF

0

0

0

MS1 ON (Select flow path 1)

1

0

0

MS2 ON (Select flow path 2)

0

1

0

MS3 ON (Select flow path 3)

0

0

1

• Output of information on flow path under measurement by multi-selector: Answer-back output

of the flow path that is being measured. ”1” when lines between terminals MSI1/MSI2 and

MSICOM are closed, and “0” when open.

Table 4.4

Output of Information on Flow Path under Measurement by Multi-selector

MSI1-MSICOM

MSI2-MSICOM

MS OFF

0

0

MS1 (Flow path 1 under

measurement)

1

0

MS2 (Flow path 2 under

measurement)

0

1

MS3 (Flow path 3 under

measurement)

1

1

CAUTION

Be careful because the contact output is open when the power of the OX400 is OFF.

!

INLET

OUTLET

N200

C

US

R

172608

MS1

RC1
RC2
RC3

DO

FAIL

MS2

MS3

MSI1

MSI2

MSI

COM

SERIAL

(RS232)

FUSE

250V T3.15A

PUMP

OFF

MS

COM

RCCOM

+

+

-

-

mA

V

Terminal screws are M3

Primary output terminal

Secondary output terminal

Range marker contact

output terminal

Contact input

(with pump “-P”

option)

G

NOT using

Power cord connector for

two-pole plug with earthing contact

Error (Fail) contact output

terminal

Alarm (DO) contactoutput terminal

(high/low oxygen concentration alarm)

Multi-selector

terminal

(with “/MS” option)

Contact output for

switching of

measurement flow path

Contact output for

measurement flow

path data

RS232 communication

connector

Figure 4.4

Wiring Terminals

4.2.3 Communication

A D-sub 9-pin connector for RS232 communication is located on the rear panel of the OX400. RS232

connection cable length must be no longer than 3 m for CE marking.