Calibration method, Procedure – Yokogawa DAQWORX User Manual

Page 179

Advertising
background image

8-6

IM WX102-01E

8.4

Calibrating the Transmission Output Module
(DA, DC only)

Calibration Method

The following two methods may be used for calibration:

• Calibrating while checking the values with instruments such as a digital multimeter.

• Calibrating by using a previously determined value.

Note

When calibrating with a previously determined value, the DA100 or DC100 may not be able to

recognize the transmission output module depending on which calibration values are set.

Procedure

Calibrate while Confirming the Output Value

Connect the transmission output terminal to be calibrated to a DMM which satisfies the

specifications below.

Specification

Recommended Device

Accuracy

HP3458A

DCV: +/–(0.01% of rdg. +10µV)
DCA: +/–(0.01% of rdg. +0.5µA)

1.

Click the connection button in the calibration screen. The system of the DA100 or

DC100 that is connected is displayed.

A[

] mark is displayed for modules that do not require calibrating.