4.19 Setting A/D Converter Integration Time, Filter,

and Scan Interval

Open the A/D Converter Integration Time, Filter, and Scan Interval setting screen (see
below) as follows:

Click the SETUP tab and select Interval from the SETUP (Basic) Setting contents box,
or select from the menubar Settings > SETUP (Basic) Settings > and then either, A/D
Converter Integration Time, Filter, Scan Interval, or File of Retransmission Output.

Select a SETUP topic

Setup tab

Select A/D converter time

Filter ON/OFF

Measurement scan interval

Set the time constant

A/D Converter Integration Time

Select the A/D converter integration time for each unit. Select it from the following:
• Auto: the frequency of 50/60 Hz is automatically switched corresponding to the power

frequency of this instrument. Note that Auto does not function for the stand-alone DC

power supply model or for extended types using a DC power supply subunit.

• 50 Hz: the integration time is set to 20 ms (50 Hz).

• 60 Hz: the integration time is set to 16.7 ms (60 Hz).

• 100 ms: the integration time is set to 100 ms (10 Hz)

• The default setting is AUTO. However, if the instrument is a stand-alone DC power

supply model, the default setting is 20 ms (50 Hz).

Filter

Set the filter to reduce normal mode noise, ON or OFF.

Scan Interval

• The duration of time (one scan) in which the measurement of all channels is carried

out, is called the scan interval.

• This interval can be set to any value from 0.5 second to 60 seconds. The shortest is

40 ch/500 ms for the stand-alone model, or 300 ch/500 ms for the expandable model
(varies with the shortest measurement period of the input module).