22

Checking that the contents of the calibration counter are correct

These seca scales are calibrated. Calibration may be performed only by authorised cen-
tres. To ensure this, the scales are fitted with a calibration counter which records any

change to data relevant to calibration. If you want to check whether the scales are prop-
erly calibrated, proceed as follows:
– If necessary, switch off the scales.

– Keep the

FUNC/SET key depressed

and start the scales.
The current contents of the calibration

counter are shown on the display for a

few seconds.

– Compare the output contents of the cal-

ibration counter with the number shown

on the calibration counter sticker.
For valid calibration, both numbers

must match. If sticker and calibration

counter do not match, then re-calibra-
tion is required. Contact your service

agent or seca Customer Service.

If a supplementary calibration is neces-
sary, then the supplementary calibration

counter sticker illustrated opposite is used

to record the calibration counter position

instead of the seca calibration sticker

shown above. This sticker is secured with

an additional seal by the person author-
ised to perform the supplementary calibra-
tion. The supplementary calibration mark

is available from seca Customer Service

under the number 14-05-01-886.

Correct weighing

– Press the Start key with no load on the

scale.

se(A

,

8.8.8.8.8.

and

0.0

appear consecu-

tively in the display. The scale is then au-
tomatically set to zero and ready to use.

– Stand on the scales and remain

motionless.

– Read the weighing result off the digital

display.

– If a weight of more than 250 kg is

placed on the scales, then

St0P

ap-

pears in the display.

– To switch off the scales, press the Start

button again.

Switching off the device

– To switch off the scales, press the Start button again.

+