increasing conductivity, the deadband will be lower than the setpoint. i.e. the
auxiliary device connected to the relay will decrease the conductivity.

a) Enter the menu by pressing CALL, proceed by pressing CALL until the RELAY

DEADBAND LED is illuminated. Now use the arrow keys to make the display
read the desired value. Press ENTER. The display will flash until ENTER is
pressed again to confirm the value.

b) Press RUN, to place the instrument on line, or press CALL for another menu

selection.


5.6 Cycle Feature for Control Relay

5.6.1

The 2200C has a cycling feature for the control relay. This feature if enabled will
cause the relays to cycle based on the on and off times selected. This useful
feature can help eliminate overshooting, saving expensive chemicals.

a) If you wish to have cycling control for the Control Relay, you must put DIP

switch NO. 1 of Bank S2 in “ON” position. (See Section 4.7). If the dip switch is
in the “OFF” position, the Relay will act in a simple ON-OFF manner.

b) The following example describes how to enter the ON and OFF times for the

Control Relay cycling control.

i) Enter the menu by pressing Call, proceed by pressing CALL until the

CYCLE LED illuminates.

ii) With the arrow buttons make the display read the number of seconds

(maximum 600) you wish the relay to be activated. Example 30. Press
ENTER. The display will flash until ENTER is pressed again to confirm
entry.

iii) Press CALL. The LED will flash. Now make the display read the

number of seconds (maximum 600) you wish the relay to be
deactivated. Example 60. Press ENTER. Press ENTER again as
above.

iv) With this configuration when the process reaches the Control Relay

Setpoint the relay will cycle continuously 30 seconds activated, 60
seconds deactivated until the Deadband Relay A setting is passed.


NOTE: As shipped from the factory the OFF time is set at 10 seconds and the

ON time is set at 5 seconds.





5.7 Alarm High

5.7.1

The instrument is fitted with a relay, which is set to activate on both high and low
alarm conditions. The deadband is factory set. To set the ALARM HIGH proceed as

follows.