b) The "ESCAPE" procedure is to be used if a unit is serviced for a new processor,

or in case of a memory loss problem, when so advised by Water Analytics
service support.

9.2.3

Before performing the "ESCAPE" procedure it is important to know that this
procedure provides the option to erase all programmed values and replace them
with the factory set default values. This means that the range selection and the
conductivity calibration must be performed after an "ESCAPE" procedure. After
that, all the control, alarm and scaling parameters mentioned above must be set to
the user values.


9.2.4

To perform the ""ESCAPE"" proceed as follows.


a) Turn off the power.

b) Press and hold the RUN button for about 3 seconds, while turning on the power.

The FAIL LED will turn red and status code 7 will be obtained to show that the
"ESCAPE" procedure was just performed.


9.2.5

After the "ESCAPE" procedure it is necessary to do the following:


a) Tune the analog outputs and the temperature output span to suit your particular

application and loop. See Sections 7.4, 7.5, and 7.6.


b) Calibrate the system, as described in Section 5.2.


c) Set up the user values for:


• Output High and Low - See Section 5.3.

• Control Relay - See Sections 5.4 and 5.5.
• Alarm Relay - See Sections 5.7 and 5.8.


d) After all the above operations are performed the FAIL LED will turn

"green” and status "0" should be obtained.

e) The unit may be tested now, using the TEST menu item (See Section

5.8.), or the built-in self-testing feature, (See Section 9.1.5.)

e) Bring the unit on line for measurement and control.













9.3 Printed Circuit Board Replacement

9.3.1

a) To replace printed circuit boards, shut off all power to the 2200C, including any

independent power to the relay contacts. Make a record of the external wiring,
then disconnect the wires. Unplug the ribbon cable connector.


b) The power supply circuit board is fastened to the back of the enclosure by four

screws, remove the screws and the board is free. Reverse the procedure to
mount a replacement board.