Model slt-5 non-contacting level & flow monitor – Aquametrix SLT-5 Ultrasonic Flow & Level Monitor User Manual

Page 4

Advertising
background image

Model SLT-5 Non-Contacting Level & Flow Monitor

Technical Data

Monitor

Electronics Enclosure

NEMA4X (IP 66) polycarbonate with clear, shatterproof cover

Accuracy

±0.25% of Range or 2 mm (0.08”) whichever is greater, Repeatability and

Linearity: ±0.1%

Displays

Numbers: 4 digit LCD; Menu/Status/Totalizer: 16 digit alphanumeric

Operating Modes

Level, Range or Open Channel Flow

Programming

3-button keypad with menu selection. Parameters are password protected

Power Input

100-130V AC 50/60 Hz, 6 watts maximum

Output

Isolated 4-20 mA, 1000 ohm load maximum, Programmable Offset

Control Relays

2 Relays (5 amp SPDT); programmable level alarm, pump control, pump

alternation, failsafe/echo-loss, air temperature alarm

Electrical Surge Protection

Sensor, 4-20 mA and AC power input

Operating Temperature

-5° to 140°F (-20° to 60°C)

Approximate Shipping Weight 10 lbs. (4.5 kg)

Sensor

Deadband (blanking)

Programmable, Minimum 12 in (305 mm)

Beam Angle

8°, Locate sensor 1 ft from tank sidewall for every 10 ft depth

Operating Frequency:

42 KHz

Exposed Materials

PVC and Teflon

Operating Temperature

-40 to 150

0

F (-40 to 65

0

C) with automatic temperature compensation

Temperature Compensation

20 psi (1.35 bar) maximum

Sensor Cable

RG62AU coaxial, 25 ft (7.6 m) standard length (See Options)

Options

Sensors

Sensors:

Sensor Cable

50 ft. (15 m) RG62AU continuous from sensor or splice up to 500 ft (150 m)

Sensor Cable Junction Box

Watertight NEMA4 steel with connection terminal strip

Power Input

200-260 VAC 50/60 Hz, 12VDC (battery power) or 24 VDC

Control Relays

4 additional (6 total), rated 5 amp SPDT

Enclosure Heater

Thermostatically controlled - for temperatures below 32°F (0°C)

Data Logger

Built-in, 2 million data point capacity with USB output

Enclosure Panel Mount

Flange assembly mounts standard enclosure

Sensor Mounting Stand

For Open Channel Flow applications

www.WaterAnalytics.net

978-749-9949

Toll free - 855-747-7623 (855-pHProbe)

Dimensions

Advertising