Status one display mode, Analog out display mode, Analog out programming – AW Gear Meters RT-Ex15-Ex User Manual

Page 15

Advertising
background image

AW-Lake Company 8809 Industrial Drive, Franksville, WI 53126  web: www.awgearmeters.com

Tel: 262-884-9800  Fax: 262-884-9810  Email: [email protected]

RT-Ex15-Ex manual rev 6.13.11.docx

Revision 3.3.10

15


When entering or editing the KFT value, the leftmost digit will blink (cursor). Select the digit to
edit using the SEL button or magnetic switch. The cursor will move to the right and will wrap
around. The decimal point can be relocated when it is selected (blinking). Use the UP or DN
button or magnetic switch to increment or decrement the variable value or move the decimal
point. Use DN to move the decimal point to the right, UP to move it to the left. When the
desired value is displayed, use the ENT button or magnetic switch to store the programmed value
and advance to the next variable.

ENG UNITS A
- The desired engineering units for display of total are programmable using up
to four letters. Example: GALS

Initial default is

PULS


NOTE: The engineering units display does not affect the scaling in any way and serves as
nothing more than a label. The total scaling is performed in the KFT calculation.

The first (left-most) character will be highlighted with the blinking cursor first. The SEL button
or magnetic switch moves the cursor and the UP or DN button or magnetic switch scrolls
through the alphabet. When the characters are all entered, use the ENT button or magnetic switch
to store the variable. The RT-Ex15 will exit the variable programming mode and return to the
selected display.
.

STATUS ONE Display Mode

The STATUS ONE screen displays the Rate A

PL

, Total A, and the analog output value in mA.

This screen is useful for set-up or troubleshooting. A typical display would be:



ANALOG OUT Display Mode

The ANALOG OUT display mode displays the analog output command value in milliamps.
NOTE: This is the commanded value, not the actual value, and under certain conditions some
difference could occur between the displayed value and the actual mA output value. The two-
digit number shown after ANALOG OUT indicates what the output represents. In this case, “00”
indicates that the output represents RATE A

GT

. See the Analog Point parameter for more

information regarding output assignment. A typical display would be:




ANALOG OUT Programming

The signal assignment, offset and scaling values for the mA output are programmable from the
ANALOG OUT display screen. The output can represent rate or total and the flow rate or total
value represented by the maximum output is programmable. Use the MODE/DN button or
magnetic switch to select the ANALOG OUT display.