INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™

Page 15 of 44

If the body is not damaged the pipe has to be cut just after the nut and newly

installed according above instructions using new ferrules. (The nut can be re-used)

Disconnected fittings can be re-installed by hand tightening the nut followed by

wrench tightening for 1/4 turn. It is recommended to perform a leak test after re-

installation.

2.4.2.

Sample supply

The sample supply line must be heat-traced and/or insulated to keep the gas above dew

point. Check your sample data for required temperature.

Sample inlet connection on the analyzer is identified with a tag-plate.

The sample connection on the analyzer is for 1/8”OD tubing. Tubing size to the process

may require a different tube size, this should be determined taking into account process

pressure, sample line length and acceptable lag time.

2.4.3.

Calibration gasses

Calibration gas composition is depending on range and process gas. COSA INSTRUMENT

can advise suitable compositions.

For a single range analyzer, 2 calibration gasses (low and high value) are recommended.

For a dual range analyzer, 3 calibration gasses (low/medium and high value) are

mandatory.