Testing the system, Smoke test, Sensitivity test – Edwards Signaling ESL 700 SERIES User Manual

Page 3: Maintenance

Advertising
background image

3

ESL 700 Series

Testing the System

After all connections are completed and the wiring is checked per NFPA

72, apply power to the system. There should not be an alarm. If there is,

power down the system and determine whether a unit is latched or if there

is a problem with the wiring. If no alarm has occurred, go to the last unit

and check the unit power with a volt meter for the specified voltage.

Testing Each Detector

1 .

Prior to testing, remove the red dust cover. The units will not work

with the dust cover in place.

2 .

Disconnect alarm notification appliances, releasing service devices,

and extinguishing systems prior to detector tests.

3 .

Be sure to reconnect all devices at the conclusion of testing.

Smoke Test

The units should be tested in place annually using one of the following

methods:

1. Use Smoke! in a can

®

and follow the directions on the can.

OR

2. Hold a smoldering punk or cotton wick close to the unit and gently

direct the smoke into the smoke entry openings for 20 seconds or until

an alarm is indicated.

The unit performs a smoke test every 9 seconds while flashing its LED. If

smoke is detected, the rate of sampling increases to every 4.5 seconds.

Excessive smoke must be detected in three consecutive tests for the alarm

to sound. BE SURE TO PROPERLY EXTINGUISH THE SMOKE

SOURCE AFTER TESTING!

If it is successful, the LED will remain lit. This is a gross, go/no-go test and

is not a reliable indication of unit sensitivity. For in-depth sensitivity

testing, see Sensitivity Test. To reset the detector, operate the system reset

switch for 2 to 3 seconds to remove power from the detectors. Control

unit alarm and all ancillary functions should be verified for a complete test

of each detector.

Sensitivity Test

1 . Hold the magnet near the integral reed switch of the unit for more

than one second (see Figure 10). The LED will flash 1 to 9 times.

2 . Count the number of LED flashes, then use the following table to

determine if any action is necessary.

Flashes

Flashes

Flashes

Flashes

Flashes

0-1

Indication:

Unservicable hardware fault.

Action:

Reset and rerun sensitivity test. If the error

persists, replace unit.

2-3

Indication:

Unit is becoming insensitive.

Action:

Clean and reset the unit. Rerun sensitivity test. If

the error persists, replace the unit.

4-7

Indication:

Unit is within normal sensitivity range.

Action:

N/A

8-9

Indication:

Unit is becoming too sensitive.

Action:

Verify the optical chamber is snapped down

securely. Clean the unit and replace the optical

chamber.

After the sequence of blinks, if the sensitivity is found to be within limits

and if all other tests pass, the detector will go into alarm until reset by the

panel. If the sensitivity is not within limits, or an unserviceable hardware

fault has been detected, the alarm LED will continue to flash once per

second until the detector is reset by the panel. If sensitivity test indicates an

unacceptable level, take action recommended above. If action does not

result in acceptable sensitivity, replace unit.

Maintenance

If a unit drifts beyond its approved sensitivity range for more than 24 hours,

or fails internal diagnostic tests, the unit automatically indicates trouble by

flashing its LED every second. This meets field sensitivity testing require-

ments without the need for external meters.

In accordance with NFPA 72, unit sensitivity should be checked within one

year of installation and every alternate year thereafter, in commercial in-

stallations, or every three years in residential sites.

The unit’s replaceable optical block chamber unsnaps for easy field cleaning

and service. Whenever the LED indicates cleaning is necessary, follow these

steps:

1 . Remove the detector cap (see Figure 8), unsnap and throw away the

optical block chamber (see Figure 9).

2. Thoroughly blow off the optical base and snap a new optical block

chamber (part #211) into place.

NOTE:

Be sure the new optical block chamber is seated all the way

down.

3. Replace the detector cap and verify sensitivity. See Sensitivity Test.

Figure 7. Removing the detector head

with the locking tab slot removed

Figure 5. Continuity switch

Continuity switch

Figure 6. Installing the detector head

Base

Head

Figure 8. Removing detector cap

Figure 9. Optical chamber

Figure 10. Sensitivity level test

SQUEEZE

HERE

Detector cap

Field replaceable

optical chamber

Detector cap

Screwdriver slot

Advertising