4.3 Basic Operation

Before connecting the power cord to an AC line outlet, make sure that the AC line voltage input switch on the rear panel of

the instrument is correctly set for the AC line voltage. After ensuring the voltage setting, set the switches and controls of the
instrument as shown below:

Item

No

Setting

Item

No

Setting

POWER

(9)

Disengage position(OFF)

SLOPE

(22) +

INTEN

(2)

Clockwise(3-o‘clock
position)

TRIG ALT

(24) Released

FOCUS

(4)

Mid-position

LEVEL LOCK

(29) Pushed in

ILLUM

(6)

Full anti-clockwise Position

HOLDOFF

(31) MIN(anti-clockwise)

(Except 622G & 635G)

TRIGGER MODE (28) AUTO

VERT MODE (39)

CH 1

HORIZ DISPLAY (38) A (623G,653G &658Gonly)

CHOP

(41)

Released

MODE

CH 2 INV

(36)

Released

TIME/DIV

(18) 0.5mSec/DIV

POSITION

(40)(37) Mid-position

SWP.UNCAL

(19) Released

VOLTS/DIV

(10)(14) 0.5V/DIV

POSITION

(34) Mid-position

VARIABLE

(13)(17) CAL(clockwise position)

x10 MAG

(33) Released

AC-DC-GND

(11)(15) GND

X-Y

(27) Released

SOURCE

(26)

Set to CH 1

COUPLING

(25)

AC

After setting the switches and controls as mentioned, connect the power cord to the AC line outlet, and then, continue as follows:

1)

Engage the POWER switch and make sure that the power LED is turned on. In about 20 seconds, a trace will appear on the
CRT screen. If no trace appears in about 60 seconds, counter check the switch and control setting.

2)

Adjust the trace to an appropriate brightness and image with the INTEN control and FOCUS control respectively.

3)

Align the trace with the horizontal center line of the graticule by adjusting the CH 1 POSITION control and TRACE
ROTATION control (adjustable by screwdriver).

4)

Connect the probe to the CH 1 INPUT terminal and apply the 2Vp-p CALIBRATOR signal to the probe tip.