En - 16, The above said anomalous cases cannot be stored – HT instruments GEO416 User Manual

Page 17

Advertising
background image

GEO416

-

GEO416GS

EN - 16

4.

When starting a measurement, if the
red cable (connected to S terminal)
and the green cable (connected to ES
terminal) are reversed
, the instrument
do not perform the test, emits a long
sound tone and displays the screen
beside

Red and green cables
reversed

Example under ρ mode

5.

When starting a measurement, if an
interfering voltage higher than 9V is
detected at the volt circuit’s input
,
the instrument do not perform the test,
emits a long sound tone and displays
the screen beside

Too high input interfering
voltage at volt circuit

Input interfering voltage
value

Example under 2W mode

6.

When starting a measurement, if an
interfering voltage higher than 9V is
detected at the ampere circuit’s
input
, the instrument do not perform the
test, emits a long sound tone and
displays the screen beside

Too high input interfering
voltage at ampere circuit

Input interfering voltage
value

Example under 2W mode

7.

When starting a measurement, if an
interfering voltage higher than 9V is
detected at both the ampere and volt
circuits’ inputs
, the instrument do not
perform the test, emits a long sound
tone and displays the screen beside

Too high input interfering
voltage at both ampere
and volt circuits

Input interfering voltage
value

Example under 2W mode

8.

If battery voltage is too low the
instrument displays the symbol of low
battery as well as the message

and no measurement is allowed. It is
however possible to carry out settings,
reading of stored data, etc

Too low power supply, low
batteries

Input interfering voltage
value and distance set
among rods

Example under ρ mode

9.

The above said anomalous cases cannot be stored