9.1. TECHNICAL FEATURES
Accuracy indicated as [%reading + (number dgts * resolution)] at 23 ± 5°C, <80%RH

3- and 2-wire earth resistance measurement - EARTH 3W and EARTH 2W

Measurement range (**)

Resolution [

]

Accuracy (*)

Reading [

] Measure

[

]

0.01

 19.99

0.08

 19.99

0.01

(2.5%rdg + 2dgt)

20.0

 199.9

20.0

 199.9

0.1

200

 1999

200

 1999

1

2.00

 19.99k

2.00

 19.99k

0.01k

20.0

 49.9k

20.0

 49.9k

0.1k

Ground resistivity measurement -

 (distance among the rods=1m)

Range (**)

Resolution [

m]

Accuracy (*)

Reading [

m] Measure

m]

0.06

 19.99

0.50

 19.99

0.01

(2.5%rdg+2dgt)

20.0

 199.9

20.0

 199.9

0.1

200

 1999

200

 1999

1

2.00

 19.99k

2.00

 19.99k

0.01k

20.0

 199.9k

20.0

 199.9k

0.1k

200

 314k

200

 314k

1k

Measuring frequency: 77.5Hz; Test current:

 12mA; Open loop voltage: < 25Vrms

Disturbance voltage on ampere and volt circuits: the measurement is taken with the stated accuracy if the interfering

voltage is

 3V, while for interfering voltages ranging from > 3V and  9V, the accuracy decreases progressively;

with an interfering voltage equal to 9V the instrument does not perform the test

(*) If

R

P

> 1200 + 100 R

X

and/or R

C

> 1200 + 100 R

X

, R

P

> 50k

 and/or R

C

> 50k

 and the instrument carries out the

test, the accuracy of the instrument is

(10% reading) where:

R

P

= resistance of the voltage circuit; R

C

= resistance of the current circuit; R

E

= measured earth resistance

(**) Automatic selection of the range


Interfering voltage measurement

Range (**)

Resolution [V]

Uncertainty

Reading [V]

Measure [V]

0

 460

7

 460

1

(2.0%rdg+2dgt)

(**) Automatic selection of the range