HT instruments GEO416 User Manual

Page 23

Advertising
background image

GEO416

-

GEO416GS

EN - 22

9.1.1. Reference

standards

Safety: IEC/EN61010-1,

IEC/EN61557-1, IEC/EN61557-5

Measuring accessories’ safety:

IEC/EN61010-031

Technical literature: IEC/EN61187
Insulation: double

insulation

Mechanical protection:

IP50 according IEC/EN60529

Pollution level:

2

Measurement category:

CAT III 240V (to earth), maximum 415V between inputs

Max height of use:

2000m (6562ft)


9.1.2. General

features

Mechanical features
Dimensions (L x W x H):

235 x 165 x 75mm (9 x 6 x 3in)

Weight (including batteries):

1kg (2lv)


Power supply
Battery type:

6x1.5V batteries type AA LR6 or

6x1.2V rechargeable batteries NI-MH AA LR6 2100mA

Low battery indication:

symbol “

” is displayed

Battery life:

approx. 500 tests

Auto Power OFF:

after 3 min of idleness


Display
Features:

LCD custom with back-light 73x65 mm


Memory
Features:

999 memory locations


PC connection
Features:

optoinsulated port for bi-directional communication


9.2. ENVIRONMENT
9.2.1. Operating

environmental

conditions

Reference calibration temperature: 23 ± 5°C (73 ± 41°F)
Working temperature:

0 ÷ 40°C (32°F ÷ 104°F)

Maximum relative humidity:

<80%RH

Storage temperature:

-10 ÷ 60°C (14°F ÷ 140°F)

Storage humidity:

<80%RH

This instrument complies with the requirements of the European Low Voltage

Directive 2006/95/CE (LVD) and EMC Directive 2004/108/CE

9.3. ACCESSORIES
9.3.1. Standard and optional accessories GEO416
See enclosed packing list

9.3.2. Standard accessories GEO416GS

 Set of 4 cables, banana-banana, L=1m

KIT416CV

 Set of 4 alligator clips

COC4-UK

Carrying

bag BORSA2000N

 ISO9000 calibration certificate

 User

manual