Caution – HT instruments HT32 User Manual

Page 12

Advertising
background image

HT32

EN - 11

4.3.3. DC Current measurement

CAUTION

Maximum input for DC current is 10A. Do not attempt to take any current
measurement exceeding such limit to avoid the risk of electrical shock and
damages to the instrument.

Fig. 4: DC Current measurement


1. Switch OFF the circuit under test.
2. Turn the selector on A.
3. Insert the red test lead into A jack and the black plug into COM jack (Fig. 4).
4. Connect red and black plugs in series with the circuit under test respecting the polarity

and current flow indicated in Fig. 4.

5. Power the circuit under test. The current value is displayed.
6. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from

display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.

7. The message "OL" means that the current exceeds the measuring limits. In this case

disconnect the test leads from the circuit under test to avoid damaging the instrument
or endanger your own safety.

8. If the measured value is lower than 6mA, to get a better resolution:

 Switch off the circuit under test.
 Turn the selector on A.
 Remove the red test lead from A jack, and insert it into

HzVA jack.

 Power the circuit under test.

9. The symbol "-" on the display means that current has opposite direction than shown in

Fig. 4.

10. For minimum and maximum value measurement and HOLD function please refer to

paragraph 4.2.