6. SETTING MEASURING FUNCTIONS

In each measuring mode (see § 7), by pressing the ENTER key, the instrument shows the
following screen:

MEASURE SET

MEASURE SET

Emissivity

Dewpoint / Wetbulb

Alarm High

Type K

Alarm Low

Laser

Auto Mode

Max/Min

Average/Diff

Ambient Temp. / %RH

Fig. 14: Measurement setting menu


Use the arrow keys

S or T to select the items and confirm with ENTER to access the

relevant programming sections.


6.1. EMISSIVITY

SETTING

The radiating power or emissivity “

ε

” is a value between 0 and 1 which indicates the

capacity of an object to emit energy in the infrared frequency range. This parameter
depends both on the material the object is made of and on the finishing (painting).
Parameter “

ε

” is set by default to 0.95 because, in practice, it has been established that

approx. 90% of the objects have such a radiating power.

1. Select the first item (

ε = x.xx) and press the ENTER key to set emissivity (see Fig. 15

and Table 3). Use the arrow keys

S or T to set the value and then the ENTER key to

confirm

2. Use the arrow keys

S or T to select one of the common materials listed below, among

the options: Concrete, Glass, Human skin, Ice/Water, Plastic, Wood

3. Press the ESC key to save and exit the function

Fig. 15: Emissivity setting menu