HT instruments HT3320 User Manual

Page 24

Advertising
background image

HT3320

EN - 23

7.6. DATALOGGER

FUNCTION

In this mode, the instrument records the IR temperature and air temperature/humidity
values with time scan and programmable alarm conditions on the measures.

1. Switch on the instrument and enter the “DATALOGGER” mode.
2. Set the desired values of the measurement parameters (see § 6)
3. Press the ENTER key to enter the programming section of the datalogger function. The

following screen appears on the instrument’s display:

Fig. 31: Datalogger function parameter setting

4. Select the parameters using the arrow keys

S or T and press ENTER. Use the arrow

keys

S or T again to set the desired values and confirm with ENTER again. The

following parameters are programmable:

PARAMETER DESCRIPTION

High

Setting of the maximum alarm threshold value within range -50°C ÷
1000°C

Low

Setting of the minimum alarm threshold value within range -50°C ÷
1000°C

Time

Setting of the time scan among consecutive recordings within the
measuring campaign in the range 1s ÷ 3600s

Color

Setting of the color of the characters shown on the display (see §
5.4)

Measure Set

Setting of the measuring parameters (see § 6)


5. Press and release the T key to start recording. The graphic screen with the real time

temperature measurement is shown at display

6. Press the ESC key to stop recording. Data is automatically saved in the instrument’s

memory (see § 8)

7. Press the ESC key to exit the function and go back to the general menu.

Advertising